Several parameters have to be taken into consideration while choosing a microfluidic tubing. Depending on your application, the choice of the right tubing for your microfluidic experiment can strongly improve the efficiency of your flow control setup and the quality of your scientific results. While discussing with microfluidic researchers, it appeared that most of them did not take the time to correctly choose the best tubing and fittings for their particular application. For that reason we decided to create this tutorial to help researchers to save valuable time when choosing the right microfluidic tubing for their experiment.
A high dead volume in a tube or connector can have 4 drawbacks:
1 The time for a full replacement of the liquid can be longer than you think, due to the parabolic profile of the liquid speed in the tube, full replacement of the liquid in the tube requires the passage of an amount of liquid which is higher than the volume of the tube. The transfer time can be approximated as follows: t=V/ φ where V is the tube dead volume and φ the flow rate. In this case, 1 volume of the tube is transferred but the composition of the liquid at the outlet still have residual amounts of the previous liquid of ~30%. The time of transfer (and liquid amount) have to be multiplied by a factor of 6 to reach an elimination of the previous liquid > 95%.
2 A high transfer time of the fluid along the tube: the time for the replacement of the liquid in a tube or the time to fill the tube while experiment has started.
3 A large consumption of sample if the tube has to be filled with an expensive or rare sample. Dead volume can be calculated as follow: V =L*π*ID²/4 where ID is the internal diameter and L the length of the tube.
4 The internal volume of the tube can change because of its deformation if the tube is moved during manipulation or just when flows are stopped because of the fluid inertia. For low flow rates application, these volume changes can induce unexpected flows or oscillations in the microfluidic chips. These artefacts strongly depend on the configuration of the setup and are difficult to predict, but in general, to limit these effects, the use of large tubing have to be avoided for low flow rate applications.
For very low flow rate applications, pressure can deform the tube, leading to a change of its internal volume. This change of volume is often responsible for slow down in pressure equilibration when dead-end tubings are used. The deformation of the tube due to internal pressure can be roughly approximated using the following relations: dV/V=2P/E (for thick walled tubes) or dV/V=P*R/(E*t) (for thin walled tubes), where P is the pressure change value, E is the Young modulus of the tube material, R is the tube radius and t the thickness of the tube (values for common tubes are given below). To limit compliance effects of tubes, small and rigid tubes are preferred and can Syringe Pump Flow Stabilizer Microfluidic Kit But sometimes the high compliance of a soft tubing can also be employed in a sophisticated manner to Syringe Pump Flow Stabilizer Microfluidic Kit often observed when using a syringe pump.
For thin capillary, the pressure drop along the tube can become non-negligible and or dominating comparing to the pressure drop in the microfluidic chip. This pressure drop, induced by the fluidic resistance of the tube can be approximated by using Poiseuille law as follow: dP =8Lη.Q/pi.R^4. Where L is the tube length, Q is the flow rate in the tube and η is the viscosity of the fluid. Typical numbers are given in the table below. Fluidic resistances are often used in microfluidic setups to get rid of flow rate instabilities and pressure variations created by droplet generation, or enable low fluidic resistance setups to work with low flow rates.
Syringe Pump Flow Stabilizer Microfluidic Kit
(1) for a 10 cm length tube, (2) for a 10 cm length tube,given per µL/min of flow rate for water, (3) volume change for a tube of 10 cm, and a pressure change of 1 bar, * Depends on connexion type.
Email* I hereby agree than Elveflow uses my personal data
Do you want tips on how to best set up your microfluidic experiment? Do you need inspiration or a different angle to take on your specific problem? Well, we probably have an application note just for you, feel free to check them out!
Free-flow electrophoresis (FFE) is a technique that enables the continuous separation of analytes as they flow through a planar channel.
The profile of laminar flow through a small straight pipe may be approximated by small concentric cylinders towards the direction of the flow.
Finding the right technique for particle encapsulation using micro and nanoparticles is key for a successful particle encapsulation protocol.
This review demonstrates the development of microfluidic techniques and their capability in performing various types of chemical synthesis.
A guide to calculate the flow resistance of your microfluidic circuit.
One of the key criteria to choose your microfluidic device material is its chemical resistance. This review will help you choose one depending on your application.
Microfluidics: definitions. This short review aims to cover the main microfluidics definitions from the simple word by word science to the advantages & applications
Introduction to thermoelectric sensor | Microfluidics immunosensors offer multiple advantages over the conventional immunoassays that include improved reaction rate, reduced time for incubation of the reactants, and decreased reagents and sample consumption. Moreover, miniaturization and integration of the multiple assay components permit automation, precise flow control, increased reproducibility, and the possibility for high-throughput analysis.
PLGA nanoparticles are of great interest for biomedical applications. This review focus on microfluidic synthesis methods of these nanoparticles.
This review focuses on point of care (POC) diagnostic devices for pathogen detection
Acoustic techniques for sorting and separation of micron-sized particles in microfluidic devices
Microfluidics is the science that deals with the flow of liquid inside micrometer-size channels.
Air bubbles are among the most recurring issues in microfluidics. Because of the micrometric dimensions of the tubes and channels
Microrheology is a technique used to measure the rheological properties of a medium, such as viscosity and viscoelasticity. This short review is presenting the different techniques and applications.
Diagnosis is the first step in treating any disease, and paper microfluidic devices can facilitate this crucial step. Most of the time, this necessary task is
Get a quote
Name*
Email*
Message
Newsletter subscription
We will answer within 24 hours
By filling in your info you accept that we use your data.
Collaborations
Need customer support?
Serial Number of your product
Support Type AdviceHardware SupportSoftware Support
Subject*
I hereby agree that Elveflow uses my personal data Newsletter subscription
How can we help you?
Message I hereby agree that Elveflow uses my personal data Newsletter subscription